Continuous glucose monitoring provides maximal information about shifting blood glucose levels throughout the day and facilitates the making of optimal treatment decisions for the diabetic patient. This report discusses continuous glucose monitoring in terms of its purposes, technologies, target populations, accuracy, clinical indications, outcomes, and problems. In this context, the medical literature on continuous glucose monitoring available through the end of 2004 is reviewed.

Continuous glucose monitoring provides information about the direction, magnitude, duration, frequency, and causes of fluctuations in blood glucose levels. Compared with conventional intensified glucose monitoring, defined as three to four blood glucose measurements per day, continuous monitoring provides much greater insight into glucose levels throughout the day. Continuous glucose readings that supply trend information can help identify and prevent unwanted periods of hypo- and hyperglycemia.

The difference between an intermittent and a continuous monitor for monitoring blood glucose is similar to that between a regular camera and a continuous security camera for monitoring an important situation. A regular camera takes discrete, accurate snapshots; its pictures do not predict the future; it produces a small set of pictures that can all be carefully studied; and effort is required to take each picture. A continuous security camera, on the other hand, takes multiple, poorly focused frames; displays a sequential array of frames whose trend predicts the future; produces too much information for each frame to be studied carefully; and operates automatically after it is turned on. The two types of blood glucose monitors differ in much the same way: 1) an intermittent blood glucose monitor measures discrete glucose levels extremely accurately, whereas a continuous monitor provides multiple glucose levels of fair accuracy; 2) with an intermittent monitor, current blood glucose levels do not predict future glucose levels, but with a continuous monitor, trends in glucose levels do have this predictive capability; 3) with an intermittent monitor, it is easy to study every measured blood glucose value over most time periods, but with a continuous monitor, too many data are generated to study all data points; and 4) an intermittent blood glucose monitor requires effort to operate, whereas a continuous monitor does not. Returning to the camera analogy, just as the best tool for closely monitoring a situation when the outcome is important often may be a continuous security camera rather than a regular camera, likewise the best way to monitor diabetes often may be a continuous glucose monitor (CGM) rather than an intermittent monitor.

Five CGMs have been approved by the U.S. Food and Drug Administration (FDA) for use in the U.S. or carry CE marking for use in Europe. They are the Continuous Glucose Monitoring System Gold (CGMS Gold; Medtronic MiniMed, Northridge, CA) (1), the GlucoWatch G2 Biographer (GW2B; Cygnus, Redwood City, CA) (2), the Guardian Telemetered Glucose Monitoring System (Medtronic MiniMed) (3), the GlucoDay (A. Menarini Diagnostics, Florence, Italy) (4), and the Pendra (Pendragon Medical, Zurich, Switzerland) (5). A sixth monitor, whose premarket approval application has been submitted to the FDA, is the FreeStyle Navigator Continuous Glucose Monitor (Abbott Laboratories, Alameda, CA) (6).

The currently available CGMs measure blood glucose either with minimal invasiveness through continuous measurement of interstitial fluid (ISF) or with the noninvasive method of applying electromagnetic radiation through the skin to blood vessels in the body. The technologies for bringing a sensor into contact with ISF include inserting an indwelling sensor subcutaneously (into the abdominal wall or arm) to measure ISF in situ or harvesting this fluid by various mechanisms that compromise the skin barrier and delivering the fluid to an external sensor (7). These ISF measurement technologies are defined as minimally invasive because they compromise the skin barrier but do not puncture any blood vessels. After a warm-up period of up to 2 h and a device-specific calibration process, each device's sensor will provide a blood glucose reading every 1–10 min for up to 72 h with the minimally invasive technology and up to 3 months with the noninvasive technology. Results are available to the patient in real time or retrospectively. Every manufacturer of a CGM produces at least one model that sounds an alarm if the glucose level falls outside of a preset euglycemic range. The available and likely soon-to-be-available CGMs are compared in Table 1.

Based on their mechanisms, specifications, and performance records, each CGM offers a particular set of features that are attractive for patients and clinicians. Table 2 presents three of these features for each available and likely soon-to-be-available CGM.

Invasive indwelling intravascular sensors that measure blood glucose directly are also under development for monitoring hospitalized patients. Prolonged use of such devices might cause vascular damage or infection. No articles have been published on their performance.

Proper patient selection will ensure safe use of CGMs. Patients should be motivated to participate in their diabetes care, informed about diabetes, and mechanically adept. These characteristics are even more necessary for real-time than for retrospective monitors. Currently available CGMs require up to four finger-stick (not alternate site) blood glucose measurements per day for calibration. The ideal time to calibrate is either after fasting or at least 3 h postprandially, but not right after exercise or when the blood glucose level is likely to be rising or falling. Without such calibration, continuous readings may be inaccurate. Currently available CGMs that provide real-time readings should not be used to make therapeutic decisions, such as whether to dose insulin or eat, because they are not sufficiently accurate. Instead, an abnormal reading should prompt a finger-stick blood glucose measurement whose value can be acted upon. Patients require a thorough training program to calibrate and operate a CGM. Furthermore, patients using real-time CGMs require extensive diabetes education to make safe and effective therapeutic decisions. In particular, they must know the duration of their boluses of insulin to avoid repeatedly and excessively injecting themselves with unnecessary extra doses of insulin over a short duration for hyperglycemia that does not immediately respond to the first insulin dose.

The second-generation CGMS Gold and its first-generation predecessor, CGMS, together have been studied in the greatest number of articles about continuous glucose monitoring in the medical literature. A CGM can be used to evaluate 24-h blood glucose profiles in diabetic patients who are receiving glucose-lowering drugs. Using Medtronic MiniMed's CGMS, Levetan et al. (8) and Abrahamian et al. (9) quantified the effect of pramlintide and nateglinide on blood glucose, respectively.

A real-time CGM can be programmed to sound an alarm for readings below or above a target range (10). The most important use of an alarm is to detect unsuspected hypoglycemia (such as during sleep) so that glucose can be administered to prevent brain damage. There is a trade-off between an alarm's sensitivity and specificity. In general, if the alarm is set to sound at a lower level than the hypoglycemic threshold, then the specificity will be good but the sensitivity may be poor. If the alarm is set to sound at a glucose level higher than the hypoglycemic threshold, then the sensitivity will be good but the specificity may be poor. The greater accuracy a continuous monitor can provide, the less of a trade-off is necessary (11).

The Diabetes Research in Children Network (DirecNet) is a U.S. network of five clinical centers and a coordinating center dedicated to researching glucose monitoring technology in children with type 1 diabetes (12). The network's investigators, the DirecNet Study Group, assessed the accuracy of the first- and second-generation CGMS and the GW2B in children with type 1 diabetes in concurrently published studies (13,14). The second-generation CGMS Gold, compared with the first-generation CGMS, had a lower median relative absolute difference (RAD) between CGMS glucose and reference serum glucose paired values (11 and 19%, respectively) (13). For the GW2B, the median RAD between GW2B glucose and reference serum glucose paired values was 16% (14). Similar RAD values of 21% have been reported for the first-generation CGMS by Kubiak et al. (15). RAD values of 12.8% (16) and 12.8–15.7% (17) have been reported for the second-generation CGMS Gold system by Goldberg et al. (16) and Guerci et al. (17), respectively.

Djakoure-Platonoff et al. (18) evaluated CGMS accuracy in 53 type 1 diabetic inpatients and outpatients who wore a CGMS for 3–6 days. The mean absolute differences between sensor and blood glucose meter values were 25–34 mg/dl.

Maran et al. (4), using the GlucoDay CGM, observed that for a mixed type 1/type 2 diabetic population, the percent bias between the GlucoDay and venous blood levels was −2.0% in the hypoglycemic range (<70 mg/dl), 6.9% in the euglycemic range (70–180 mg/dl), and 11.2% in the hyperglycemic range (>180 mg/dl). Poscia and colleagues found the GlucoDay RAD to be 20% in rabbits (19) and 15% in a mixed type 1/type 2 diabetic population (20).

Bode et al. (3) evaluated the performance of the Guardian Continuous Monitoring System (Medtronic MiniMed) and whether using real-time alarms reduced hypo- and hyperglycemic excursions in a type 1 diabetic population. The mean absolute relative error between home blood glucose meter readings and sensor values was 21.3% (median 7.3%); further, on average the Guardian read 12.8 mg/dl below the concurrent home blood glucose meter readings. The hypoglycemia alert was able to distinguish glucose values ≤70 mg/dl with 67% sensitivity, 90% specificity, and 47% false alerts. The hyperglycemia alert showed a similar ability, detecting values ≥250 mg/dl with 63% sensitivity, 97% specificity, and 19% false alerts. The alerts resulted in a significant (P = 0.03) reduction in the duration of hypoglycemic excursions and a marginally significant (P = 0.07) increase in the frequency of hyperglycemic excursions.

The DirecNet Study Group (21) found the CGMS Gold system, which is the second generation of CGMS technology, as well as the GW2B, which is the second generation of GlucoWatch technology, to have inversely proportional sensitivity and specificity rates during hypoglycemia in children and adolescents with type 1 diabetes. A series of alarm settings were selected for a reference blood glucose of 60 mg/dl. For CGMS Gold, the settings with sensitivity and specificity were 60 mg/dl, 49 and 42%; 80 mg/dl, 84 and 36%; 100 mg/dl, 100 and 25%; and 120 mg/dl, 100 and 16%. With the GW2B, the settings were 60 mg/dl, 23 and 49%; 80 mg/dl, 59 and 33%; 100 mg/dl, 84 and 20%; and 120 mg/dl, 92 and 15%. The authors concluded, "These data show that the GW2B and the CGMS do not reliably detect hypoglycemia. Both of these devices perform better at higher glucose levels, suggesting that they may be more useful in reducing HbA1c levels than in detecting hypoglycemia" (21).

Tsalikian et al. (22) observed the sensitivity and specificity of the hypoglycemic alarm and the down alert alarm on the GW2B in 89 children and adolescents with type 1 diabetes. The "down alert" alarm is triggered when extrapolation of the current glucose trend anticipates that hypoglycemia will occur within the next 20 min. Overnight sensitivity was 23% with the hypoglycemia alarm alone and 77% when combined with the down alert alarm. For the hypoglycemia alarm alone, the false-positive rate was 16%, but for the two alarm types combined, the false-positive rate increased to 62%. The down alert alarm improved the sensitivity of the GW2B to detect hypoglycemia but resulted in a large increase in the false alarm rate.

Conrad et al. (23) evaluated the sensitivity and specificity of the CGMS for reference capillary blood glucose levels <60 mg/dl in five children and infants with hypoglycemic disorders. They reported a sensitivity of 65%, a specificity of 91%, and a false-positive rate of 43%.

The International Organization for Standardization (ISO) standards for accuracy of point blood glucose tests require that a sensor blood glucose value be within 15 mg/dl of reference for a reference value ≤75 mg/dl and within 20% of reference for a reference value >75 mg/dl. Sensor accuracy by this definition is expressed as the percentage of data pairs meeting these requirements (24). The DirecNet group found that for hypoglycemic blood glucose levels (determined by a reference blood glucose monitor, the OneTouch Ultra), the CGMS Gold met the ISO standards in only 48% of readings and the GW2B met these standards in only 32% of readings (13,14). The percentage of data points attaining ISO accuracy standards climbed as the blood glucose level rose, topping out for the highest segment of reference blood glucose levels (i.e., blood glucose values ≥240 mg/dl). In this glycemic category, the CGMS Gold and GW2B, respectively, met ISO accuracy for 81 and 67% of data points (13,14). In a separate series of 15 healthy nondiabetic children undergoing continuous glucose monitoring over 24 h, the DirecNet Group reported that the median absolute difference in concentrations for the GW2B was 13 mg/dl and for the CGMS was 17 mg/dl. Furthermore, 30% of the values from the GW2B and 42% of the values from the CGMS deviated by >20 mg/dl from the reference value (25).

The DirecNet investigators also noted that the second-generation CGMS (CGMS Gold) devices that they tested performed better than first-generation CGMS devices (21). The second-generation algorithm had previously been reported to be more useful for avoiding false hypoglycemic and hyperglycemic readings in nondiabetic children, whereas both algorithms gave similar results in diabetic children (26). The overreading of nocturnal hypoglycemia with the first-generation CGMS devices had been suspected because of both a high incidence of asymptomatic nocturnal hypoglycemia reported by the CGMS (without simultaneous blood glucose monitor readings) in children with type 1 diabetes (27–29) and documented spurious hypoglycemic readings (30,31). The sensor malfunction rate of 18–28% (31–33) in the first-generation CGMS sensors has been improved in the second-generation CGMS Gold system; a malfunction rate of 17–18% has been reported in two series of CGMS Gold users (16–17).

Kovatchev et al. (34) have proposed a continuous glucose error grid analysis (EGA) system for analyzing the performance of CGMs. This tool is an updated version of the EGA that was developed to assess the performance of intermittent blood glucose monitors (35). The continuous glucose EGA combines a point-EGA (p-EGA; corrected for time lags between fluctuations in ISF and blood during dynamic fluctuations in circulating glucose levels) with a rate EGA (r-EGA). The continuous glucose EGA classifies readings of combined p-EGA and r-EGA as "accurate," "benign," or "erroneous." The use of r-EGA fine-tunes the assessment of accuracy compared with the use of p-EGA only. The continuous glucose EGA can judge the performance of a CGM separately at low, normal, and high blood glucose levels.

The mean blood glucose level measured by a CGMS over 72 h correlates, to some extent, with a patient's HbA1c level, which reflects the mean blood glucose level over the previous 1–4 months (36). A correlation between mean sensor glucose value, determined by CGMS, and the HbA1c value has been observed to be r = 0.53 (P = 0.002) (37) and r = 0.59 (P = 0.002) (38).

Situations that require detailed information about blood glucose fluctuations that only continuous monitoring can provide include when adjusting therapy, quantifying the response in a trial of a diabetes therapy, assessing the impact of lifestyle modifications on glycemic control, monitoring conditions where tighter control without hypoglycemia is sought (e.g., gestational diabetes, pediatric diabetes, in the intensive care unit), diagnosing and then preventing hypoglycemia (e.g., during sleep, with hypoglycemia unawareness), and diagnosing and preventing postprandial hypoglycemia. The most important use of continuous blood glucose monitoring is to facilitate adjustments in therapy to improve control. Specific therapeutic adjustments include changing from regular to a synthetic ultrashort-acting insulin analog at mealtime, changing from NPH to a synthetic ultralong-acting insulin once or twice per day, increasing or decreasing the mealtime insulin bolus dosage, increasing or decreasing the basal insulin rate, altering the treatment of intermittent hypoglycemia or hyperglycemia, changing the insulin-to-glucose correction algorithm for premeal hyperglycemia, changing the insulin-to-carbohydrate ratio at mealtime, changing the method for counting carbohydrates, changing the carbohydrate composition of the diet, changing the discount in short-acting insulin dosage for exercise, changing the nighttime regimen because of the dawn phenomenon, changing the target preprandial or postprandial blood glucose values, or before referring a patient for psychological counseling to improve adherence to the treatment regimen.

The most frequent therapy adjustment by Sabbah et al. (39) (out of eight adjustments) was to increase the mealtime bolus dosage. The most frequent therapy adjustment by Kaufman et al. (40) (out of nine adjustments) was to modify the type of basal long-acting insulin.

Of five randomized controlled studies of CGM technologies that have used HbA1c as a surrogate marker for morbidity and mortality related to diabetes (27,41–44) four evaluated the CGMS (27,42–44) and one evaluated the GW2B (41). In each study, the use of continuous glucose monitoring compared with standard monitoring was associated with improved mean HbA1c levels.

Chase et al. (27) in 2001 reported a 1-month randomized, controlled pilot trial of the CGMS in 11 children as a supplement to standard capillary blood glucose monitoring to improve HbA1c. The CGMS group experienced a statistically significant decrease in mean HbA1c of 0.36% (P < 0.01), whereas the control subjects experienced a nonsignificant decrease in HbA1c of 0.2%. The difference between changes in the two groups was 0.16%, which is not a very important difference in such a small study. At 3 months (2 months postintervention), four of the five children who used the CGMS continued to have lower (but no longer significantly lower) HbA1c values in comparison with their initial values (mean decrease for the group of five CGMS users = 1.04%; P = 0.07); three of the six control participants also had lower HbA1c values at 3 months (mean decrease for the group of six control subjects = 0.62%).

In a 2003 study, Chase et al. (41) reported that compared with standard therapy, the GlucoWatch Biographer significantly improved glucose control. They randomly assigned 40 children to diabetes management with or without the GlucoWatch Biographer. Both groups performed conventional blood glucose monitoring four times per day. Subjects in the Biographer group used the Biographer for 3 months (intervention phase) and were then followed for an additional 6 months (observational phase). Biographer wearers were asked to check their blood glucose reading if the Biographer glucose reading was ≤70 or ≥300 mg/dl. After 3 months, HbA1c values improved from 8.9 at baseline to 8.4% in Biographer users and worsened from 8.6 to 9.0% in control subjects. The difference between Biographer users and control subjects (8.4 vs. 9.0%; P < 0.05) at 3 months was statistically significant. During the observation phase following the intervention phase, the HbA1c in the Biographer group remained lower than in the control group (at months 6 and 9 of the study), but the differences were no longer statistically significant.

Chico et al. (42) in 2003 reported on the results of a 3-month randomized, controlled trial of 75 type 1 diabetic subjects who were treated with the CGMS (40 subjects in the treatment group) or intensive capillary glucose measurement (35 subjects in the control group) (42). HbA1c concentrations decreased significantly in both the CGMS (from 8.3 to 7.5%; P < 0.01) and the control (from 8.0 to 7.5%; P < 0.01) groups. However, the CGMS did not result in statistically better outcomes compared with capillary glucose measurements.

In 2003, Ludvigsson and Hanas (43) reported on the results of a controlled, crossover trial comparing the effect of using a CGMS or seven-point glucose profiles on HbA1c. During the open arm of the trial, the 27 type 1 diabetic subjects wore the CGMS for 3 days every 2 weeks for 3 months, and during the blinded (to CGMS data) arm, the subjects checked seven-point glucose profiles every week for 3 months. At 3 months, the two study arms were crossed over. The HbA1c levels decreased significantly in the open arm using the CGMS (from 7.70 to 7.31%; P = 0.013), but not in the blinded arm (7.75 to 7.65%; NS); the difference between study arms was significant (7.31 vs. 7.65%; P = 0.011).

In 2004, Tanenberg et al. (44) reported on a 12-week trial comparing CGMS with standard glucose monitoring in 128 insulin-treated diabetic subjects. In both groups, the HbA1c levels decreased significantly compared with baseline values, but there was a nonsignificant improvement in HbA1c outcomes in the CGMS group compared with the standard monitoring group. HbA1c levels decreased by 0.8% in the CGMS group and 0.7% in the standard group (P = 0.70). The CGMS group, however, had a significantly shorter mean duration of hypoglycemic events (sensor glucose ≤ 60 mg/dl) at week 12 of the study (49 vs. 81 min; P = 0.009).

An additional five nonrandomized, uncontrolled trials of continuous glucose monitoring have demonstrated a statistically significant improvement in HbA1c using this technology in addition to usual capillary blood glucose monitoring (37,40,45–47). The studies are discussed below.

In 1999, Bode et al. (45) reported the first pilot study of the clinical performance of the CGMS in an uncontrolled, nonrandomized trial of nine type 1 diabetic subjects. After a pair of 1-week courses of monitoring followed by therapeutic adjustments based on CGMS readings, mean HbA1c levels fell from 9.9 at baseline to 8.8% (P = 0.0006) within 5 weeks of the trial's commencement. The improved HbA1c levels were sustained: 10 weeks into the study, the mean HbA1c of this population had further declined to 8.6% (P = 0.019) (48).

Kaufman et al. (40) reported in 2001 on an uncontrolled, nonrandomized study of the effect on HbA1c of a single 72-h course of CGMS monitoring in a cohort of 47 pediatric subjects. The continuous information was used to alter therapy. Mean HbA1c levels fell from 8.6 to 8.4 and 8.3% (both paired Student's t test, P ≤ 0.03) 3 and 6 months, respectively, after the course of CGMS compared with at the time of placement.

Schiaffini et al. (46) reported in 2002 the findings of a study in which the CGMS was applied on two occasions 6 weeks apart for 72 h per each application. The study was initially conducted on 27 type 1 diabetic children, all of whom received an initial course of CGMS monitoring and 18 of whom elected to continue in the 6-week study to receive therapy adjustments and a second course of CGMS monitoring upon completion of the study. All continuing subjects self-monitored capillary blood glucose levels 4–5 times daily. Insulin therapy was adjusted based on the initial CGMS results and subsequent spot capillary glucose levels. Of the 18 subjects who fully participated in the study, compared with the single CGMS cohort, there was a significant fall in mean serum fructosamine levels from 349 to 330 μmol/l (P < 0.05). Study participants completing the study also experienced a decreased number of hypoglycemic events per 72 h compared with the incidence rate among baseline measurements of those not completing the study (2.5 vs. 3.9 episodes per 72 h; P < 0.05).

In 2002, Salardi et al. (37) reported on a study of the prolonged effects of a 72-h course of CGMS monitoring in a nonrandomized, uncontrolled trial with 28 type 1 diabetic subjects. The subjects' HbA1c was lower by 0.40 and 0.43% 3 and 6 months, respectively, after the course of CGMS monitoring compared with baseline levels (37).

In 2003, Schaepelynck-Belicar et al. (47) reported the findings of a nonrandomized, uncontrolled trial of a 72-h course of CGMS monitoring. Continuous data were used to determine rational adjustments in insulin therapy in 12 type 1 diabetic subjects. Changes involved alterations of the dosage in three subjects, insulin type in seven subjects, the number of daily injections in five subjects, and the delivery technology (from insulin injection to pump therapy) in one subject. Reassessment 2 months later demonstrated a significant reduction of glycemic excursions in eight subjects and a decrease in the mean HbA1c from 10.3 to 8.75% (P < 0.05). The calculated low blood glucose index increased, but not significantly (49).

Continuous glucose monitoring technology has been used as an educational tool to document the incidence and magnitude of hypoglycemia in diabetic adults (50–52) and children (22,23,28,29,53,54), pregnant women with diabetes (55–59), patients after pancreas (60) and islet cell (61–62) transplant, and children with glycogen storage disease (63,64). This technology has similarly been used in hyperglycemia in type 1 diabetic subjects using insulin pump therapy (65,66), type 1 diabetic subjects using insulin injections (67), pregnant women with (55–59,68) and without (68,69) diabetes, gastroparesis patients (70), and cystic fibrosis patients (71). Continuous glucose monitoring has also been used as a therapeutic tool to decrease the incidence and magnitude of hypoglycemia in three studies (46,72,73).

In 2004, Weintrob et al. (72) reported the results of a study that used two 72-h courses of CGMS, 2.5 months apart, to provide information for improving glycemic patterns in a cohort of 23 type 1 diabetic children. Compared with the first CGMS reading, the second reading demonstrated a smaller 24-h area under the curve for hypoglycemia (P = 0.04), a shorter duration of nocturnal hypoglycemia (P = 0.05), and a tendency for a lower rate of diurnal hypoglycemic events (P = 0.1).

In a nonrandomized, uncontrolled multicenter series of 15 type 1 diabetic subjects who were monitored using a long-term, investigational, subcutaneously implanted continuous glucose sensor, Garg et al. (73) reported reduced hyperglycemic and hypoglycemic excursions. A control period (mean 50 days) of being blinded to real-time blood glucose data was followed by a study period (mean 44 days) of access to these data. The subjects spent a median of 47% less time under 60 mg/dl (P < 0.05) and 25% less time over 240 mg/dl (P < 0.05) during the nonblinded study period compared with during the blinded control period.

Currently available CGMs can present problems for users. One major problem is their lack of accuracy for each single data point compared with the accuracy of simultaneous intermittent blood glucose measurements. CGMs are generally least accurate in the hypoglycemic range (74). Because of the accuracy issue, it is necessary to incorporate trend information in using continuous blood glucose data. All real-time CGMs incorporate an arrow that indicates an upward or downward trend. The trend arrow can be set to display minor or major levels of upward or downward trending. A minimally invasive CGM can be associated with side effects related to chronic ISF harvesting. The GW2B has been associated with skin irritation, but this problem can be reduced without affecting the measurement accuracy by pretreating the forearm with a corticosteroid spray (75). The Pendra, which resembles the GW2B, does not harvest fluid from the skin or cause skin irritation. The CGMS rarely causes local discomfort from the implanted catheter; the sensor must be inserted at least 2 in from an insulin infusion site and 3 in from an insulin injection site.

Most currently available outpatient continuous blood glucose monitors do not actually measure the glucose concentration within whole blood, but instead measure the glucose concentration within the ISF compartment. The Pendra measures blood glucose both noninvasively and continuously. Depending on activity or feeding schedules (and especially during periods of rapid blood glucose shifting), equilibration between shifting blood and interstitial fluid glucose levels may lag (76). It is unknown whether sensor site selection can minimize this intercompartmental lag or the lag that can occur during equilibration between arterial and skin capillary blood glucose levels at sites other than the fingertip.

Reimbursement for CGMs by insurance or government payer organizations has been limited. Insurance companies are demanding rigorous scientific evidence about continuous monitoring before they will pay for this technology (77). Their stated reason is that an insufficient number of randomized, controlled trials of these devices has shown a statistically significant decrease in diabetic complications or HbA1c levels, which are accepted as a surrogate marker of long-term blood glucose control (78). Indeed, only five articles have been published describing randomized, controlled trials of continuous glucose monitoring technologies for improving clinical outcomes (27,41–44). In all five studies, continuous monitoring compared with standard monitoring was associated with improved HbA1c levels, but the improvements in HbA1c levels in the continuously monitored subjects were statistically significant in only two (41,43) of these five studies.

Continuous glucose monitoring offers the capability of expressing the frequency and severity of hypoglycemic episodes much more clearly than does intermittent glucose testing. Continuous glucose monitoring also offers the capability of expressing the mean blood glucose value in new ways. The mean blood glucose value can shift quickly with any new treatment, and it is not always practical to wait for months or weeks, respectively, for the HbA1c or fructosamine values to shift. Continuous glucose monitoring can document the time spent in the normal, low, and high ranges, which may be more valuable than a single integrated data point, such as HbA1c or fructosamine. Long-term exposure to midrange glycemia may turn out to be better for avoiding complications than exposure to many upward hyperglycemic spikes and downward hypoglycemic spikes; however, the spikes may cancel out each other in terms of altering the long-term markers. Continuous glucose monitoring can distinguish the two exposures, but long-term markers cannot stratify time spent above and below a particular target. It is possible that a new measure of glycemia, derived from the duration of normal, low, and high readings, could supplement HbA1c as an integrated measure of control. Furthermore, measurements of mean amplitude of glycemic excursions (54), composite hypoglycemic score (62), and lability index (62) could provide information about the tendency for a mean blood glucose level to be comprised of stable or labile data points. For some patients, a decreased amount of glycemic instability alone, even without any improvement in HbA1c, might represent an improved outcome.

Continuous glucose monitoring offers advantages over intermittent glucose monitoring when glycemic patterns are poorly understood. The information about direction, magnitude, duration, frequency, and causes of fluctuations in blood glucose levels that can be obtained by continuous glucose monitoring is simply not available with intermittent blood glucose monitoring. When retrospective patterns are needed to adjust therapy or document the state of physiology, CGMs are useful. When real-time recognition of both the absolute magnitude of glycemia as well as trend patterns are needed, then a real-time CGM provides a wealth of information. Technologies for continuous glucose monitoring require patient education for proper use. During hypoglycemia or periods of rapid fluctuation, values provided by CGMs may be inaccurate. Clinical outcome studies suggest that measures of mean glycemia and hypoglycemic burden both improve with the use of continuous glucose monitoring, but more studies are needed to convince payors to reimburse for this technology. In this data-hungry world, it appears likely that CGMs will eventually become a routine part of diabetes management, initially for patients with difficult-to-control diabetes and eventually for most patients with diabetes. Retrospective reporting will eventually give way to real-time readings, and adjunctive use requiring a confirmatory finger-stick blood test will eventually give way to primary use without the requirement of such confirmation. As methods for minimally invasive and noninvasive continuous monitoring advance, diabetic patients will use this technology more routinely. Data printouts from CGMs will increasingly provide a roadmap for effective diabetes management in the 21st century.

After this article was submitted for publication, an additional study was published describing the results of a multicenter, randomized, controlled trial of continuous glucose monitoring (79). In 2005, the DirecNet Study Group reported the results of a 6-month trial comparing GW2B use with standard glucose monitoring in 200 type 1 diabetic subjects ages 7–18 years. Use of the GW2B compared with standard glucose monitoring resulted in no significant difference in glycemic control, but was associated with a nonsignificant trend (P = 0.10) toward a greater number of episodes of severe hypoglycemia. Furthermore, six additional articles were also published, following the submission of this article, describing the performance of the CGMS in type 1 diabetes (80), type 2 diabetes (81), pregnancy (82), and both type 1 and type 2 diabetic patients receiving peritoneal dialysis (83); the GlucoDay in type 1 diabetes (84); and an investigational viscometric affinity sensor in type 1 diabetes (85).

Table 1—

Specifications of available and likely soon-to-be-available products for continuous glucose monitoring

Product FDA approved CE marked Year first approved or marked Sensor type Sensor mechanism Sensor location Sensor warmup (h) Calibrations per lifetime of sensor Sensor lifespan (h) Frequency of testing (min) Time of blood glucose data display Alarm
Continuous Glucose Monitoring System Gold Yes Yes 1999 Minimally invasive Enzyme-tipped catheter Subcutaneous abdomen 2 12 72 5 Retrospective No
GlucoWatch G2 Biographer Yes Yes 2001 Minimally invasive Reverse iontophoresis External on arm or forearm 2 1 13 10 Real time Yes
Guardian Telemetered Glucose Monitoring System Yes Yes 2004 Minimally invasive Enzyme-tipped catheter Subcutaneous arm 2 12 72 5 Retrospective Yes
GlucoDay No Yes 2001 Minimally invasive Microdialysis Subcutaneous abdomen 0 1 48 3 Real time or retrospective Yes
Pendra No Yes 2004 Noninvasive Impedance spectroscopy External on wrist 1 20 3 months 1 Real time Yes
FreeStyle Navigator Continuous Glucose Monitor No No Minimally invasive Enzyme-tipped catheter Subcutaneous arm 1 1 72 1 Real time Yes
Product FDA approved CE marked Year first approved or marked Sensor type Sensor mechanism Sensor location Sensor warmup (h) Calibrations per lifetime of sensor Sensor lifespan (h) Frequency of testing (min) Time of blood glucose data display Alarm
Continuous Glucose Monitoring System Gold Yes Yes 1999 Minimally invasive Enzyme-tipped catheter Subcutaneous abdomen 2 12 72 5 Retrospective No
GlucoWatch G2 Biographer Yes Yes 2001 Minimally invasive Reverse iontophoresis External on arm or forearm 2 1 13 10 Real time Yes
Guardian Telemetered Glucose Monitoring System Yes Yes 2004 Minimally invasive Enzyme-tipped catheter Subcutaneous arm 2 12 72 5 Retrospective Yes
GlucoDay No Yes 2001 Minimally invasive Microdialysis Subcutaneous abdomen 0 1 48 3 Real time or retrospective Yes
Pendra No Yes 2004 Noninvasive Impedance spectroscopy External on wrist 1 20 3 months 1 Real time Yes
FreeStyle Navigator Continuous Glucose Monitor No No Minimally invasive Enzyme-tipped catheter Subcutaneous arm 1 1 72 1 Real time Yes

Table 2—

Attractive features of available and likely soon-to-be-available continuous glucose monitors

Model Features
Continuous Glucose Monitoring System Gold Long sensor life Avoids arm implantation First product on the market and used in the most studies
GlucoWatch G2 Biographer Needle-free Real-time readings Alarm for out-of-range values
Guardian Telemetered Glucose Monitoring System Long sensor life Alarm for out-of-range values Avoids arm implantation
GlucoDay Real-time readings Infrequent calibrations Choice of retrospective or real-time data
Pendra Noninvasive No skin irritation Measures glucose in blood and not in interstitial fluid
FreeStyle Navigator Continuous Glucose Monitor Long sensor life Alarm for out-of-range values Avoids abdominal wall implantation
Model Features
Continuous Glucose Monitoring System Gold Long sensor life Avoids arm implantation First product on the market and used in the most studies
GlucoWatch G2 Biographer Needle-free Real-time readings Alarm for out-of-range values
Guardian Telemetered Glucose Monitoring System Long sensor life Alarm for out-of-range values Avoids arm implantation
GlucoDay Real-time readings Infrequent calibrations Choice of retrospective or real-time data
Pendra Noninvasive No skin irritation Measures glucose in blood and not in interstitial fluid
FreeStyle Navigator Continuous Glucose Monitor Long sensor life Alarm for out-of-range values Avoids abdominal wall implantation

1.

Gross TM, Bode BW, Einhorn D, Kayne DM, Reed JH, White NH, Mastrototaro JJ: Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use.

Diabetes Technol Ther

2

:

49

–56,

2000

2.

Potts RO, Tamada JA, Tierney MJ: Glucose monitoring by reverse iontophoresis.

Diabetes Metab Res Rev

18 (Suppl. 1)

:

S49

–S53, 2002

3.

Bode B, Gross K, Rikalo N, Schwartz S, Wahl T, Page C, Gross T, Mastrototaro J: Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the guardian continuous monitoring system.

Diabetes Technol Ther

6

:

105

–113,

2004

4.

Maran A, Crepaldi C, Tiengo A, Grassi G, Vitali E, Pagano G, Bistoni S, Calabrese G, Santeusanio F, Leonetti F, Ribaudo M, Di Mario U, Annuzzi G, Genovese S, Riccardi G, Previti M, Cucinotta D, Giorgino F, Bellomo A, Giorgino R, Poscia A, Varalli M: Continuous subcutaneous glucose monitoring in diabetic patients: a multicenter analysis.

Diabetes Care

25

:

347

–352,

2002

5.

Pfuetzner A, Caduff A, Larbig M, Schrepfer T, Forst T: Impact of posture and fixation technique on impedance spectroscopy used for continuous and non-invasive glucose monitoring.

Diabetes Technol Ther

6

:

435

–441,

2004

6.

Feldman B, Brazg R, Schwartz S, Weinstein R: A continuous glucose sensor based on wired enzyme technology: results from a 3-day trial in patients with type 1 diabetes.

Diabetes Technol Ther

5

:

769

–779,

2003

7.

Choleau C, Klein JC, Reach G, Aussedat B, Demaria-Pesce V, Wilson GS, Gifford R, Ward WK: Calibration of a subcutaneous amperometric glucose sensor. Part 1. Effect of measurement uncertainties on the determination of sensor sensitivity and background current.

Biosens Bioelectron

17

:

641

–646,

2002

8.

Levetan C, Want LL, Weyer C, Strobel SA, Crean J, Wang Y, Maggs DG, Kolterman OG, Chandran M, Mudaliar SR, Henry RR: Impact of pramlintide on glucose fluctuations and postprandial glucose, glucagon, and triglyceride excursions among patients with type 1 diabetes intensively treated with insulin pumps.

Diabetes Care

26

:

1

–8,

2003

9.

Abrahamian H, Francesconi M, Loiskandl A, Dzien A, Prager R, Weitgasser R: Evaluation of a new insulinotropic agent by using an innovative technology: efficacy and safety of nateglinide determined by continuous glucose monitoring.

Diabetes Technol Ther

6

:

31

–37,

2004

10.

Pitzer KR, Desai S, Dunn T, Edelman S, Jayalakshmi Y, Kennedy J, Tamada JA, Potts RO: Detection of hypoglycemia with the GlucoWatch Biographer.

Diabetes Care

24

:

881

–885,

2001

11.

Reach G: Which threshold to detect hypoglycemia? Value of receiver-operator curve analysis to find a compromise between sensitivity and specificity.

Diabetes Care

24

:

803

–804,

2001

12.

Weinzimer SA, Tamborlane WV, Chase HP, Garg SK: Continuous glucose monitoring in type 1 diabetes.

Curr Diab Rep

4

:

95

–100,

2004

13.

Diabetes Research in Children Network (DirecNet) Study Group: The accuracy of the GlucoWatch G2 Biographer in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Diabetes Technol Ther

5

:

791

–800,

2003

14.

Diabetes Research in Children Network (DirecNet) Study Group: The accuracy of the CGMS in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Diabetes Technol Ther

5

:

781

–789,

2003

15.

Kubiak T, Hermanns N, Schreckling HJ, Kulzer B, Haak T: Assessment of hypoglycaemia awareness using continuous glucose monitoring.

Diabet Med

2

:

487

–490,

2004

16.

Goldberg PA, Siegel MD, Russell RR, Sherwin RS, Halickman JI, Cooper DA, Dziura JD, Inzucchi SE: Experience with the Continuous Glucose Monitoring System in a medical intensive care unit.

Diabetes Technol Ther

6

:

339

–347,

2004

17.

Guerci B, Floriot M, Bohme P, Durain D, Benichou M, Jellimann S, Drouin P: Clinical performance of CGMS in type 1 diabetic patients treated by continuous subcutaneous insulin infusion using insulin analogs.

Diabetes Care

26

:

582

–589,

2003

18.

Djakoure-Platonoff C, Radermercker R, Reach G, Slama G, Selam JI: Accuracy of the continuous glucose monitoring system in inpatient and outpatient conditions.

Diabete Metab

29

:

159

–162,

2003

19.

Poscia A, Mascini M, Moscone D, Luzzana M, Caramenti G, Cremonesi P, Valgimigli F, Bongiovanni C, Varalli M: A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 1).

Biosens Bioelectron

18

:

891

–898,

2003

20.

Varalli M, Marelli G, Maran A, Bistoni S, Luzzana M, Cremonesi P, Caramenti G, Valgimigli F, Poscia A: A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 2).

Biosens Bioelectron

18

:

899

–905,

2003

21.

Diabetes Research in Children Network (DirecNet) Study Group: Accuracy of the GlucoWatch G2 Biographer and the continuous glucose monitoring system during hypoglycemia: experience of the Diabetes Research in Children Network.

Diabetes Care

27

:

722

–726,

2004

22.

Tsalikian E, Kollman C, Mauras N, Weinzimer S, Buckingham B, Xing D, Beck R, Ruedy K, Tamborlane W, Fiallo-Scharer R, Diabetes Research in Children Network (DirecNet) Study Group: GlucoWatch G2 Biographer alarm reliability during hypoglycemia in children.

Diabetes Technol Ther

6

:

559

–566,

2004

23.

Conrad SC, Mastrototaro JJ, Gitelman SE: The use of a continuous glucose monitoring system in hypoglycemic disorders.

J Pediatr Endocrinol Metab

17

:

281

–288,

2004

24.

International Organisation for Standardisation:

Requirements for In Vitro Blood Glucose Monitoring Systems for Self-Testing in Managing Diabetes Mellitus.

ISO/TC 212/WG3. Draft International Standard ISO/DIS 15197. Geneva, International Organisation for Standardisation,

2001

25.

The Diabetes Research in Children Network (DirecNet) Study Group: Lack of accuracy of continuous glucose sensors in healthy, nondiabetic children: results of the Diabetes Research in Children Network (DirecNet) accuracy study.

J Pediatr

144

:

770

–775,

2004

26.

Weinzimer SA, DeLucia MC, Boland EA, Steffen A, Tamborlane WV: Analysis of continuous glucose monitoring data from non-diabetic and diabetic children: a tale of two algorithms.

Diabetes Technol Ther

5

:

375

–380,

2003

27.

Chase HP, Kim LM, Owen SL, MacKenzie TA, Klingensmith GJ, Murtfeldt R, Garg SK: Continuous subcutaneous glucose monitoring in children with type 1 diabetes.

Pediatrics

107

:

222

–226,

2001

28.

Boland E, Monsod T, Delucia M, Brandt CA, Fernando S, Tamborlane WV: Limitations of conventional methods of self-monitoring of blood glucose: lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes.

Diabetes Care

24

:

1858

–1862,

2001

29.

Kaufman FR, Austin J, Neinstein A, Jeng L, Halvorson M, Devoe DJ, Pitukcheewanont P: Nocturnal hypoglycemia detected with the continuous glucose monitoring system in pediatric patients with type 1 diabetes.

J Pediatr

141

:

625

–630,

2002

30.

McGowan K, Thomas W, Moran A: Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes.

Diabetes Care

25

:

1499

–1503,

2002

31.

Metzger M, Leibowitz G, Wainstein J, Glaser B, Raz I: Reproducibility of glucose measurements using the glucose sensor.

Diabetes Care

25

:

1185

–1191,

2002

32.

Amin R, Ross K, Acerini CL, Edge JA, Warner J, Dunger DB: Hypoglycemia prevalence in prepubertal children with type 1 diabetes on standard insulin regimen: use of continuous glucose monitoring system.

Diabetes Care

26

:

662

–667,

2003

33.

Sachedina N, Pickup JC: Performance assessment of the Medtronic-MiniMed Continuous Glucose Monitoring System and its use for measurement of glycaemic control in type 1 diabetic subjects.

Diabet Med

20

:

1012

–1015,

2003

34.

Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL: Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data.

Diabetes Care

27

:

1922

–1928,

2004

35.

Clarke WL, Cox DJ, Gonder-Frederick LA, Pohl SL: Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Diabetes Care

10

:

622

–628,

1987

36.

Miedema K: Laboratory tests in diagnosis and management of diabetes mellitus: practical considerations.

Clin Chem Lab Med

41

:

1259

–1265,

2003

37.

Salardi S, Zucchini S, Santoni R, Ragni L, Gualandi S, Cicognani A, Cacciari E: The glucose area under the profiles obtained with continuous glucose monitoring system relationships with HbAlc in pediatric type 1 diabetic patients.

Diabetes Care

25

:

1840

–1844,

2002

38.

Sharp P, Rainbow S: Continuous glucose monitoring and haemoglobin A(1c).

Ann Clin Biochem

39

:

516

–517,

2002

39.

Sabbah H, Mcculloch K, Fredrickson LP, Gross TM, Bode BW: Detailed data from a Continuous Glucose Monitoring System (CGMS) facilitates comprehensive diabetes management recommendations (Abstract).

Diabetes

49 (Suppl. 1)

:

A393

,

2000

40.

Kaufman FR, Gibson LC, Halvorson M, Carpenter S, Fisher LK, Pitukcheewanont P: A pilot study of the continuous glucose monitoring system: clinical decisions and glycemic control after its use in pediatric type 1 diabetic subjects.

Diabetes Care

24

:

2030

–2034,

2001

41.

Chase HP, Roberts MD, Wightman C, Klingensmith G, Garg SK, Van Wyhe M, Desai S, Harper W, Lopatin M, Bartkowiak M, Tamada J, Eastman RC: Use of the GlucoWatch Biographer in children with type 1 diabetes.

Pediatrics

111

:

790

–794,

2003

42.

Chico A, Vidal-Rios P, Subira M, Novials A: The continuous glucose monitoring system is useful for detecting unrecognized hypoglycemias in patients with type 1 and type 2 diabetes but is not better than frequent capillary glucose measurements for improving metabolic control.

Diabetes Care

26

:

1153

–1157,

2003

43.

Ludvigsson J, Hanas R: Continuous subcutaneous glucose monitoring improved metabolic control in pediatric patients with type 1 diabetes: a controlled crossover study.

Pediatrics

111

:

933

–938,

2003

44.

Tanenberg R, Bode B, Lane W, Levetan C, Mestman J, Harmel AP, Tobian J, Gross T, Mastrototaro J: Use of the continuous glucose monitoring system to guide therapy in patients with insulin-treated diabetes: a randomized controlled trial.

Mayo Clin Proc

79

:

1521

–1526,

2004

45.

Bode BW, Gross TM, Thornton KR, Mastrototaro JJ: Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study.

Diabetes Res Clin Pract

46

:

183

–190,

1999

46.

Schiaffini R, Ciampalini P, Fierabracci A, Spera S, Borrelli P, Bottazzo GF, Crino A: The Continuous Glucose Monitoring System (CGMS) in type 1 diabetic children is the way to reduce hypoglycemic risk.

Diabetes Metab Res Rev

18

:

324

–329,

2002

47.

Schaepelynck-Belicar P, Vague P, Simonin G, Lassmann-Vague V: Improved metabolic control in diabetes adolescents using the continuous glucose monitoring system (CGMS).

Diabete Metab

29

:

608

–612,

2003

48.

Gross TM, Mastrototaro JJ: Efficacy and reliability of the continuous glucose monitoring system.

Diabetes Technol Ther

2(Suppl. 1)

:

S19

–S26, 2000

49.

Kovatchev BP, Cox DJ, Gonder-Frederick LA, Young-Hyman D, Schlundt D, Clarke W: Assessment of risk for severe hypoglycemia among adults with IDDM: validation of the low blood glucose index.

Diabetes Care

21

:

1870

–1875,

1998

50.

Cheyne E, Kerr D: Making 'sense' of diabetes: using a continuous glucose sensor in clinical practice.

Diabetes Metab Res Rev

18 (Suppl. 1)

:

S43

–S38, 2002

51.

Hay LC, Wilmshurst EG, Fulcher G: Unrecognized hypo- and hyperglycemia in well-controlled patients with type 2 diabetes mellitus: the results of continuous glucose monitoring.

Diabetes Technol Ther

5

:

19

–26,

2003

52.

Cameron FJ, Ambler GR: Does continuous glucose monitoring have clinical utility in contemporary management of diabetes?

J Paediatr Child Health

40

:

79

–84,

2004

53.

Jeha GS, Karaviti LP, Anderson B, Smith EO, Donaldson S, McGirk TS, Haymond MW: Continuous glucose monitoring and the reality of metabolic control in preschool children with type 1 diabetes.

Diabetes Care

27

:

2881

–2886,

2004

54.

Alemzadeh R, Loppnow C, Parton E, Kirby M: Glucose sensor evaluation of glycemic instability in pediatric type 1 diabetes mellitus.

Diabetes Technol Ther

5

:

167

–173,

2003

55.

Chen R, Yogev Y, Ben-Haroush A, Jovanovic L, Hod M, Phillip M: Continuous glucose monitoring for the evaluation and improved control of gestational diabetes mellitus.

J Matern Fetal Neonatal Med

14

:

256

–260,

2003

56.

Yogev Y, Ben-Haroush A, Chen R, Kaplan B, Phillip M, Hod M: Continuous glucose monitoring for treatment adjustment in diabetic pregnancies: a pilot study.

Diabet Med

20

:

558

–562,

2003

57.

Yogev Y, Chen R, Ben-Haroush A, Phillip M, Jovanovic L, Hod M: Continuous glucose monitoring for the evaluation of gravid women with type 1 diabetes mellitus.

Obstet Gynecol

101

:

633

–638,

2003

58.

Yogev Y, Ben-Haroush A, Chen R, Rosenn B, Hod M, Langer O: Undiagnosed asymptomatic hypoglycemia: diet, insulin, and glyburide for gestational diabetic pregnancy.

Obstet Gynecol

104

:

88

–93,

2004

59.

Kerssen A, de Valk HW, Visser GH: The Continuous Glucose Monitoring System during pregnancy of women with type 1 diabetes mellitus: accuracy assessment.

Diabetes Technol Ther

6

:

645

–651,

2004

60.

Esmatjes E, Flores L, Vidal M, Rodriguez L, Cortes A, Almirall L, Ricart MJ, Gomis R: Hypoglycaemia after pancreas transplantation: usefulness of a continuous glucose monitoring system.

Clin Transplant

17

:

534

–538,

2003

61.

Kessler L, Passemard R, Oberholzer J, Benhamou PY, Bucher P, Toso C, Meyer P, Penfornis A, Badet L, Wolf P, Colin C, Morel P, Pinget M, GRAGIL Group: Reduction of blood glucose variability in type 1 diabetic patients treated by pancreatic islet transplantation: interest of continuous glucose monitoring.

Diabetes Care

25

:

2256

–2262,

2002

62.

Ryan EA, Shandro T, Green K, Paty BW, Senior PA, Bigam D, Shapiro AM, Vantyghem MC: Assessment of the severity of hypoglycemia and glycemic lability in type 1 diabetic subjects undergoing islet transplantation.

Diabetes

53

:

955

–962,

2004

63.

Hershkovitz E, Rachmel A, Ben-Zaken H, Phillip M: Continuous glucose monitoring in children with glycogen storage disease type I.

J Inherit Metab Dis

24

:

863

–869,

2001

64.

Maran A, Crepaldi C, Avogaro A, Catuogno S, Burlina A, Poscia A, Tiengo A: Continuous glucose monitoring in conditions other than diabetes.

Diabete Metab Res Rev

20 (Suppl. 2)

:

S50

–S55, 2004

65.

Deiss D, Hartmann R, Hoeffe J, Kordonouri O: Assessment of glycemic control by continuous glucose monitoring system in 50 children with type 1 diabetes starting on insulin pump therapy.

Pediatr Diabetes

5

:

117

–121,

2004

66.

Mlcak P, Fialova J, Trnkova K, Chlup R: A continuous glucose monitoring system (CGMS): a promising approach for improving metabolic control in persons with type 1 diabetes mellitus treated by insulin pumps.

Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub

148

:

33

–38,

2004

67.

Manuel-y-Keenoy B, Vertommen J, Abrams P, Van Gaal L, De Leeuw I, Messeri D, Poscia A: Postprandial glucose monitoring in type 1 diabetes mellitus: use of a continuous subcutaneous monitoring device.

Diabetes Metab Res Rev

20 (Suppl. 2)

:

S24

–S31, 2004

68.

Buhling KJ, Kurzidim B, Wolf C, Wohlfarth K, Mahmoudi M, Wascher C, Siebert G, Dudenhausen JW: Introductory experience with the continuous glucose monitoring system (CGMS; Medtronic Minimed) in detecting hyperglycemia by comparing the self-monitoring of blood glucose (SMBG) in non-pregnant women and in pregnant women with impaired glucose tolerance and gestational diabetes.

Exp Clin Endocrinol Diabetes

112

:

556

–560,

2004

69.

Yogev Y, Ben-Hroush A, Chen R, Rosenn B, Hod M, Langer O: Diurnal glycemic profile in obese and normal weight nondiabetic pregnant women.

Am J Obstet Gynecol

191

:

949

–953,

2004

70.

Tanenberg RJ, Pfeifer MA: Continuous glucose monitoring system: a new approach to the diagnosis of diabetic gastroparesis.

Diabetes Technol Ther

2 (Suppl. 1)

:

S73

–S80, 2000

71.

Dobson L, Sheldon CD, Hattersley AT: Conventional measures underestimate glycaemia in cystic fibrosis patients.

Diabet Med

21

:

691

–696,

2004

72.

Weintrob N, Schechter A, Benzxzaquen H, Shalitin S, Lilos P, Galatzer A, Phillip M: Glycemic patterns detected by continuous subcutaneous glucose sensing in children and adolescents with type 1 diabetes mellitus treated by multiple daily injections vs continuous subcutaneous insulin infusion.

Arch Pediatr Adolesc Med

158

:

677

–684,

2004

73.

Garg SK, Schwartz S, Edelman SV: Improved glucose excursions using an implantable real-time continuous glucose sensor in adults with type 1 diabetes.

Diabetes Care

27

:

734

–738,

2004

74.

Klonoff DC: The need for separate performance goals for glucose sensors in the hypoglycemic, normoglycemic, and hyperglycemic ranges.

Diabetes Care

27

:

834

–836,

2004

75.

Tamada JA, Davis TL, Leptien AD, Lee J, Wang B, Lopatin M, Wei C, Wilson D, Comyns K, Eastman RC: The effect of preapplication of corticosteroids on skin irritation and performance of the GlucoWatch G2 Biographer.

Diabetes Technol Ther

6

:

357

–367,

2004

76.

Kulcu E, Tamada JA, Reach G, Potts RO, Lesho MJ: Physiological differences between interstitial glucose and blood glu-cose measured in human subjects.

Diabetes Care

26

:

2405

–2409,

2003

77.

Flamm CR: Use of intermittent or continuous interstitial glucose monitoring in patients with diabetes mellitus. Technology Evaluation Center. Blue Cross and Blue Shield Association Assessment Program.

Vol. 18

, No. 16. December

2003

78.

Goldstein DE, Little RR, Lorenz RA, Malone JI, Nathan D, Peterson CM: Tests of glycemia in diabetes.

Diabetes Care

18

:

896

–909,

1995

79.

The Diabetes Research in Children Network (DirecNet) Study Group: A randomized multicenter trial comparing the GlucoWatch Biographer with standard glucose monitoring in children with type 1 diabetes.

Diabetes Care

28

:

1101

–1106,

2005

80.

Larsen J, Ford T, Lyden E, Colling C, Mack-Shipman L, Lane J: What is hypoglycemia in patients with well-controlled type 1 diabetes treated by subcutaneous insulin pump with use of the continuous glucose monitoring system?

Endocr Pract

10

:

324

–329,

2004

81.

Nyback-Nakell A, von Heijne M, Adamson U, Lins PE, Landstedt-Hallin L: Accuracy of continuous nocturnal glucose screening after 48 and 72 hours in type 2 diabetes patients on combined oral and insulin gherapy.

Diabetes Metab

30

:

517

–521,

2004

82.

Buhling KJ, Kurzidim B, Wolf C, Wohlfarth K, Mahmoudi M, Wascher C, Siebert G, Dudenhausen JW: Introductory experience with the continuous glucose monitoring system (CGMS; Medtronic Minimed) in detecting hyperglycemia by comparing the self-monitoring of blood glucose (SMBG) in non-pregnant women and in pregnant women with impaired glucose tolerance and gestational diabetes.

Exp Clin Endocrinol Diabetes

112

:

556

–560,

2004

83.

Schwing WD, Erhard P, Newman LN, Nodge MM, Czechanski BJ, Orlin SM, Walden SM, Behm K, Cacho CP, Negrea LA, Siu DS, Kern EO, Weiss MF: Assessing 24-hour blood glucose patterns in diabetic patients treated by peritoneal dialysis.

Adv Perit Dial

20

:

213

–216,

2004

84.

DeVries JH, Wentholt IM, Masurel N, Mantel I, Poscia A, Maran A, Heine RJ: Nocturnal hypoglycaemia in type 1 diabetes—consequences and assessment.

Diabetes Metab Res Rev

20 (Suppl. 2)

:

S43

–S46,

2004

85.

Diem P, Kalt L, Haueter U, Krinelke L, Fajfr R, Reihl B, Beyer U: Clinical performance of a continuous viscometric affinity sensor for glucose.

Diabetes Technol Ther

6

:

790

–799,

2004

A table elsewhere in this issue shows conventional and Système International (SI) units and conversion factors for many substances.